Fit the probe snuggly into the ear canal, then push
and release the Start button
.
Select age with Age Precision
™
button as shown below.
Press button to toggle through each age group.
NOTE: You must select an age in order to take a measurement.
The lens lter indicator will blink to signal one is
needed.
To achieve accurate measurements, make sure a new,
clean lens lter is in place before each measurement.
Attach a new lens lter by pushing the thermometer
probe straight into the lens lter inside the box and
then pulling out.
Note: The Braun ThermoScan will not work unless a
lens lter is attached.
ExacTemp light will pulse while temperature measurement is in progress. The
light will remain solid for 3 seconds to indicate that a successful temperature
reading has been achieved.
NOTE: If the probe has been properly inserted into the ear canal during the
measurement, a long beep will sound to signal the completed measurement.
If the probe has NOT been constantly placed in a stable position in the ear
canal, a sequence of short beeps will sound, the ExacTemp light will go out
and the display will show an error message (POS = position error).
See «Errors and troubleshooting» section for more
information.
6
7
9
6520
6020
The con rmation beep indicates that an accurate temperature measurement has been taken.
The result is shown on the display.
Temperature displays.
0-3 3-36 36+
Months
0-3 3-36 36+
Months
0-3 3-36 36+
Months
Age range
0-3 Months
3-36 Months
36 Months -
adult
Green
Normal temperature
≥35.8 – ≤37.4 °C
(≥96.4 – ≤99.4 °F)
≥35.4 – ≤37.6 °C
(≥95.7 – ≤99.6 °F)
≥35.4 – ≤37.7 °C
(≥95.7 – ≤99.9 °F)
Yellow
Elevated temperature
>37.6 – ≤38.5 °C
(>99.6 – ≤101.3 °F)
>37.7 – ≤39.4 °C
(>99.9 – ≤103.0 °F)
Red
High temperature
>37.4 °C
(>99.4 °F)
>38.5 °C
(>101.3 °F)
>39.4 °C
(>103.0 °F)
For the next measurement, press Eject button to
remove and discard used lens lter, and put on a new,
clean lens lter.
NOTE: Defaults to last age setting used if you do not
change.
The Braun ThermoScan ear thermometer turns o
automatically after 60 seconds of inactivity.
The thermometer can also be turned o by
pressing the Power button
.
The display will brie y ash OFF and it will
go blank.
8
Temperature-taking hints
Always replace disposable lens lters to maintain accuracy and hygiene. The right ear
measurement may di er from the measurement taken in the left ear. Therefore, always take the
temperature in the same ear. The ear must be free from obstructions or excess earwax build-up to
take an accurate reading.
External factors may in uence ear temperatures, including when an individual has:
Factor A ects
Poor probe placement
Used lens lter
Dirty lens
English
60206520
The thermometer is ready to take a temperature when the display looks like the images
below.